User's Manual

Title and text Explanation Next steps
Replace battery now
Delivery stopped. Battery
must be replaced to
resume delivery.
Insulin delivery has
stopped due to low
power. Battery was
not replaced after
the Low battery
Pump alert.
Replace the battery immediately to
resume insulin delivery. For details,
see Removing the battery, on
page 23.
Reservoir estimate at 0 U
To ensure insulin delivery,
change reservoir.
Your reservoir level
is estimated at
0 units.
Select OK to clear the alert.
Change the reservoir now.
Resume bolus?
XXX of YYY U delivered.
Resume delivery of
ZZZ U?
A normal bolus
delivery has been
interrupted because
the pump battery
was removed. If it is
within 10 minutes
since this
interruption, you
can resume this
bolus.
Check the message to see how
much of the bolus was actually
delivered.
To cancel remaining amount of
bolus, select Cancel.
To resume remaining amount of
bolus, select Resume.
Resume Dual bolus?
XX of YY U delivered.
Resume delivery of ZZ U
for XX:XX hr?
The Square portion
of Dual Bolus
delivery has been
interrupted. If it is
within 10 minutes
since this
interruption, you
can resume this
bolus.
Check the message to see how
much of the Dual Wave bolus
was actually delivered.
To cancel remaining amount of
bolus, select Cancel.
To resume remaining amount of
bolus, select Resume.
214 Chapter 12
MP6025958-022 / A
RELEASED