User's Manual

Title and text Explanation Next steps
Pump error
Delivery stopped. Current
settings cleared. Pump
restart needed. Select OK
to restart and then re-
enter your settings. See
User Guide.
Your pump
encountered an
error and will
restart. Your pump
settings will return
to factory default
values.
When the pump restarts, follow
instructions on the pump display.
After restart, check settings and
re-enter values as needed.
If you recently saved backup
settings in Manage Settings, use
Restore Settings.
•If the pump was delivering a
bolus or filling the cannula, check
Daily History and reevaluate your
need for insulin.
If this alarm recurs frequently,
write down the error code
displayed on the alarm screen
(you can also find it in your
Alarm History) and contact your
local representative for assistance
with your pump.
Pump error
Delivery stopped. Settings
unchanged. Pump restart
needed. Select OK to
restart. See User Guide.
A pump error has
occurred, you need
to restart your
pump.
Select OK to restart your pump.
•If the pump was delivering a
bolus or filling the cannula, check
Daily History and reevaluate your
need for insulin.
If this alarm recurs frequently,
write down the error code
displayed on the alarm screen
(you can also find it in your
Alarm History) and contact your
local representative for assistance
with your pump.
212 Chapter 12
MP6025958-022 / A