User's Manual

Title and text Explanation Next steps
Max Fill reached
4X.X U. Remove reservoir
and select Rewind to
restart New Reservoir
procedure.
You have exceeded
the number of units
expected to fill the
tubing. By now,
insulin should be at
the end of the
tubing.
Remove the reservoir.
Check if you still have insulin in
the reservoir. If you do, you can
continue using the same
reservoir.
Select Rewind to restart the new
reservoir procedure.
No reservoir detected
Rewind before loading
reservoir.
There is no reservoir
in the pump or the
reservoir is not
properly locked into
place.
Select Rewind.
Ensure that your reservoir is filled
with insulin.
When prompted, ensure that
your reservoir is inserted and
properly locked into place.
Power error detected
Delivery stopped. Record
your settings by
uploading to CareLink or
write your settings on
paper. See User Guide.
The internal power
source in your
pump is unable to
charge. Your pump
is operating on the
AA battery only.
Select OK to clear the alert.
Check your BG and treat as
necessary.
Record your settings as soon as
possible because your AA battery
may not last long.
Call your local representative for
assistance with your pump.
Power loss
AA battery was removed
for more than 10 min or
power was lost. Select OK
to re-enter time and date.
Your pump battery
has been out for
more than
ten minutes, and
your pump has lost
power. You must
reset your time and
date.
Select OK to go to the Time &
Date screen.
Enter the current time, time
format, and date.
alarms, alerts, and messages
Alarms, alerts, and messages 211
MP6025958-022 / A