Low reservoir
XX hours remaining.
Change reservoir.
or:
XX units remaining.
Change reservoir.
Your reservoir is low
on insulin,
according to the
number of hours or
units set in the Low
Reservoir Reminder.
• Select OK to clear the alert.
• Change the reservoir soon.
• If you do not change the
reservoir after you receive this
alert, you will receive a second
Low reservoir alert when the
insulin level reaches half of your
original alert amount. For more
details, see Low Reservoir
reminder, on page 132.
Manage settings error
Delivery stopped. Backup
settings cleared from
Manage Settings. Current
settings are working
properly. Select OK to
restart. See User Guide.
A pump error has
occurred, and you
need to restart your
pump. Your backup
settings have been
lost, but your
current settings are
unchanged.
• Select OK to restart your pump.
Your current settings are
unchanged. Only your backup
settings are lost.
• When the pump restarts, follow
instructions on the pump display.
•If the pump was delivering a
bolus or filling the cannula, check
Daily History and evaluate your
need for insulin.
• Consider saving your current
settings. For details, see Saving
your settings, on page 142.
Max Fill reached
3X.X U. Did you see drops
at the end of tubing?
You have exceeded
the number of units
expected to fill the
tubing. By now,
insulin should be at
the end of the
tubing.
• If you see drops at the end of
the tubing, select Yes.
• If you do not see drops, select
No.
• Follow instructions displayed on
the pump.