User's Manual
Title and text Explanation Next steps
Insulin flow blocked
Fill Tubing stopped.
Remove reservoir and
select Rewind to restart.
Your pump has
detected the insulin
flow was blocked
while filling the
tubing. Possible
connection issue
between the tubing
and reservoir.
• Remove the reservoir and select
Rewind to restart the fill tubing
process.
• Disconnect tubing from reservoir.
• Be sure tubing is not crimped or
bent.
•Continue following the steps
displayed on the pump using the
same infusion set and reservoir.
• If this alarm occurs again, use a
new infusion set.
If a fill tubing delivery was in
progress when the alarm occurred:
• Check the Daily History screen for
the amount of bolus already
delivered before the pump
alarmed.
• Consider delivering remaining
bolus, if the bolus insulin was not
included in an insulin injection.
Loading incomplete
Remove reservoir and
select Rewind to restart
loading.
You pressed
after loading began.
• Remove the reservoir to start
again.
• Select Rewind and follow the on-
screen instructions.
Low battery Pump
Replace battery soon.
The battery in the
pump is low on
power.
• Select OK to clear the alert.
• Replace the AA battery as soon
as possible. Otherwise, insulin
delivery stops, and the Replace
Battery Now alarm occurs.
•If the pump is delivering a bolus
or filling the cannula, wait until
delivery is complete to replace
battery.
■
alarms, alerts, and messages
Alarms, alerts, and messages 209
MP6025958-022 / A
RELEASED