User's Manual
SG has been rising at a rate of 2 mg/dL per minute or more, but less than
3 mg/dL per minute.
SG has been falling at a rate of 2 mg/dL per minute or more, but less than
3 mg/dL per minute.
SG has been rising at a rate of 3 mg/dL per minute or more.
SG has been falling at a rate of 3 mg/dL per minute or more.
Silencing Glucose Alerts
The Alert Silence feature allows you to make sensor glucose alerts silent for a set
period of time. This is useful in situations where you do not want to disturb others,
such as when you are in a business meeting or in a movie theater. When using
this feature, your system still records the time and glucose value for any alerts that
occur. You can view this information in the Alarm History screen. See Alarm History,
on page 122 for details.
If a glucose alert occurs when you are using the Alert Silence feature, the
notification light begins to flash and the Sensor alert occurred message appears
letting you know an alert was silenced, but there is no vibration or beep. If you
have not cleared the alert by the end of the preset alert silence duration, your
pump begins to beep or vibrate periodically until the alert is cleared.
The following table describes the glucose alerts that are silenced with each option.
This Alert
Silence setting
Silences these alerts
High Alerts Only Alert on high, Alert before high, and Rise Alert
High & Low
Alerts
Alert on high, Alert before high, Rise Alert, Alert on low, Alert
before low, Suspend before low, and Resume Basal Alert
■