User's Manual

You can calibrate anytime. However, calibrating with two or three down trend
arrows may temporarily decrease accuracy until the next calibration. For an
example of trend arrows on the Home screen, see Home screen with CGM, on
page 152.
If you are not using a compatible Bayer meter, enter your BG meter reading
into the pump immediately after testing your BG. Your BG meter reading is
only valid for 12 minutes, do not wait to enter it later.
Always use clean, dry fingers when you test your blood glucose levels.
Use only your fingertips when obtaining blood samples for calibration.
Note: If your BG meter readings are significantly different than your
sensor glucose readings, you need to wash your hands and calibrate
again.
Disconnecting the transmitter from the sensor
Always refer to your transmitter user guide for instructions on disconnecting the
transmitter from the sensor.
Removing the sensor
Always refer to the sensor user guide for instructions on how to remove the
sensor.
Turning off Sensor Settings
You can turn off Sensor Settings at any time. If you disconnect the transmitter
from the sensor, turn off the Sensor Settings to avoid getting a sensor alert. Your
sensor settings remain in your pump. You cannot make changes to the settings
until you turn on the Sensor Settings again.
To turn off Sensor Settings:
1. Go to the Sensor Settings screen.
Menu > Sensor Settings
2. Select Sensor.
3. Select Yes to turn off the sensor feature.
setting up continuous glucose monitoring
Setting up Continuous Glucose Monitoring 187
MP6025958-022 / A
RELEASED