User's Manual
After you have entered your first two calibrations, you must calibrate your sensor
again within 12 hours. If you do not enter a BG meter reading within 12 hours,
your pump displays the Calibrate now alert and stops calculating sensor glucose
values until a calibration BG is successfully entered. The sensor must be calibrated
at a minimum of every 12 hours throughout the life of the sensor.
If you are using a compatible Bayer meter, you can set up your pump to calibrate
automatically with each valid BG meter reading. For more details, see Setting up
Auto Calibration, on page 184.
Note: Sensor calibration is successful only if your BG entry is in the range
of 40 to 400 mg/dL. Remember to calibrate three to four times
throughout the day for optimal results.
To calibrate your sensor:
1. Take a BG meter reading.
2. Go to the Calibrate Sensor screen.
Menu > Sensor Settings > Calibrate Sensor
3. Select BG and enter the value.
4. Select Calibrate.
You can set up a reminder to notify you when your next calibration is due.
For more information, see Calibration reminders, on page 133.
Setting up Auto Calibration
The Auto Calibration feature determines how to calibrate the pump when using a
wirelessly linked compatible Bayer meter. When the Auto Calibration feature is
turned on, the pump uses any BG value in the range of 40 to 400 mg/dL for
calibration. When this feature is turned off, the pump asks you every time if you
want to use the compatible Bayer meter for calibration.
To turn on Auto Calibration:
1. Go to the Auto Calibration screen.
Menu > Sensor Settings > Auto Calibration
2. Select Auto Calibration to turn on the feature.
184 Chapter 10
MP6025958-022 / A
RELEASED