User's Manual
4. Use the pump navigation buttons to enter the serial number of the
transmitter and select OK.
The New Device screen appears.
5. Select Search on your pump and immediately remove the transmitter from
the charger.
The following happens when you start the search process:
• On your pump, a message appears to let you know your pump is
searching.
• On your transmitter, a green light flashes briefly and then turns off.
Note: The search process can take up to two minutes. You cannot
access your pump screens or suspend your pump during the search
process.
6. Place the transmitter next to the pump.
If the connection is successful, your pump displays a success message. If the
Sensor feature is turned on, the Connection icon
appears on the status
bar.
180 Chapter 10
MP6025958-022 / A
RELEASED