User's Manual

c. Select Suspend on low to have insulin suspended when you reach or fall
below your low limit. The Alert on low alert is automatically turned on
and cannot be turned off.
d. Select Alert on low if you want to receive an alert when your SG reaches
or falls below your low limit. If either suspend feature is on, this will
already be on.
e. Select Resume basal alert if you want an alert when basal insulin
delivery is resumed based on SG values during a suspend by sensor
event. If you do not turn on the alert, the Basal delivery resumed
message will still appear on the pump, but you will not be alerted.
Note: When setting your low alerts:
If you turn on the Suspend before low or the
Suspend on low feature, then the Alert on low
feature is turned on automatically.
Only one suspend by sensor feature can be used
during each time segment. You cannot use both
the Suspend before low and the Suspend on low
features in the same time segment.
8. If you entered an End time of anything other than 00:00, another time
segment appears.
When you are done entering your low settings, select Done.
9. Review your settings, and select Save.
To make changes to your Low Settings:
1. Go to the Low Settings screen.
Menu > Sensor Settings > Low Settings
The Low Settings screen appears.
2. Select Setup.
3. Select Edit.
4. Select, and if needed, adjust the time segment you would like to change.
setting up continuous glucose monitoring
Setting up Continuous Glucose Monitoring 173
MP6025958-022 / A
RELEASED