User's Manual
WARNING: Your insulin is suspended for up to two minutes while
your pump runs the self test.
Self Test includes the following tests:
Test Description
Display Turns on the display for up to 30 seconds.
Notification light Turns on the notification light for three seconds and then turns it
off.
Vibration Generates two vibration cycles.
Tone Generates an alert tone, an Easy Bolus (step 1) tone, and an alarm
tone.
The pump will run through a series of tests as listed in the previous table. Self Test
requires you to observe the pump during the test.
To run the Self Test:
1. Go to the Self Test screen.
Menu > Utilities > Self Test
A message indicates that the Self Test is in progress.
Self Test takes up to two minutes to complete. During that time, the display
briefly turns white, the notification light blinks, the pump vibrates, and the
pump beeps.
2. If Self Test does not detect a problem, the display returns to the Utilities
screen.
If Self Test detects a problem, a message appears with more information
about the problem. If Self Test displays an error message or you observe the
pump not behaving as indicated during the test, contact the 24 Hour
HelpLine or your local representative.
Sensor Demo
Sensor Demo lets you see what the Home screen would look like if you were
using the optional Continuous Glucose Monitoring (CGM) feature. For more
information about sensor graphs, please see The sensor graph, on page 191.