for an alert when your SG has been rising at a rate of
2 mg/dL per minute or more.
•
Select
for an alert when your SG has been rising at a rate of
3 mg/dL per minute or more.
Select OK, and skip to step 11.
Note: These arrows appear on your Home screen to let you know
the rate at which your SG has been rising.
10. To enter a custom rise limit, do the following:
a. Select Custom. The Custom Limit screen appears.
b. Select Rise Limit and set a rise rate from 1 to 5 mg/dL/min. You set the
rate in 0.1 mg/dL/min increments.
c. Select OK to return to the Rise Limit screen and then select OK again to
confirm your settings.
11. When you have set all the high settings for the selected time segment, select
Next to continue.
12. If you entered an End time of anything other than 00:00, another time
segment appears. When you are finished entering high settings, select Done.
13. Review your settings, and select Save.
To change your High Settings:
1. Go to the High Settings screen.
Menu > Sensor Settings > High Settings
The High Settings screen appears.
2. Select Setup.
3. Select Edit.
4. Select and, if needed, adjust the time segment you would like to change.
5. Select any alert setting to turn it on or off or to adjust the setting.
6. Select Next.
7. Select Done.