User's Manual

Manually resume your basal insulin delivery, the SmartGuard suspend by
sensor features will be unavailable for 30 minutes after your basal insulin
delivery resumes.
If your pump has been suspended for two hours and you have not responded,
basal insulin delivery automatically resumes.
If you respond within 30 minutes of basal insulin delivery being resumed, the
SmartGuard suspend by sensor features will be unavailable for a total of 30
minutes. For example:
If you respond 10 minutes after your basal insulin delivery resumes, the
SmartGuard suspend by sensor features will be unavailable for an additional
20 minutes.
If you respond 20 minutes after your basal insulin delivery resumes, the
SmartGuard suspend by sensor features will be unavailable for an additional
10 minutes.
If you respond 30 minutes to four hours after your basal insulin delivery resumes,
the SmartGuard suspend by sensor features will be available immediately.
If you do not respond, the SmartGuard suspend by sensor features will be
unavailable for four hours after basal delivery resumes.
Alert on low
The Alert on low feature is automatically turned on when either the Suspend
before low feature or the Suspend on low feature is turned on.
When Alert on low is set to on, you receive an alert when your SG reading reaches
or falls below your low limit. If your pump is suspended and you have not
responded, an emergency message appears.
Automatically resuming basal delivery after a SmartGuard suspend by sensor
event
In addition to suspending insulin delivery, the pump can also automatically resume
delivery of basal insulin. If insulin has been suspended by either the Suspend
before low or the Suspend on low feature, insulin delivery will automatically be
resumed if either of the following conditions are met:
setting up continuous glucose monitoring
Setting up Continuous Glucose Monitoring 163