User's Manual

High glucose
setting
Description
Rise Alert The Rise Alert notifies you when your glucose is rising rapidly. This
alert helps you understand how much your glucose levels are affected
by meals or, for example, when forgetting to give a bolus.
Rise Limit The Rise Limit determines when you will receive a Rise Alert. Rise Limit
is only available when using Rise Alert.
To set up your high settings, see Setting up the High Settings, on page 167.
Low settings
The low settings allow you to be alerted and/or have insulin delivery suspended
when you are either approaching or have reached your low limit. This is done by
using alerts and the SmartGuard feature described on page 152.
The following graph shows the different low settings you can use:
Low Limit
Alert before low
Alert on low
Suspend before low
Suspend on low
WARNING: SmartGuard suspend by sensor features are not
intended to be a treatment for low blood glucose. Having insulin
suspended when glucose is low may not bring your blood glucose
back to your target range for several hours. Always confirm your
blood glucose readings with your BG meter and treat according to
the recommendations of your healthcare professional.
The following sections describe the SmartGuard feature and the low settings. For
details on setting up SmartGuard and your low settings, see Setting up the Low
Settings, on page 171.