User's Manual

WARNING: Make sure the infusion set is disconnected from your
body before you remove the reservoir from your pump.
To remove your reservoir:
1. Wash your hands.
2. Remove the entire infusion set from your body.
3. If you have the optional activity guard attached to the reservoir compartment
on your pump, remove it now.
4. Turn the tubing connector a half-turn counter-clockwise, then pull the
reservoir and connector out from the pump.
tubing connector
5. Dispose of the used reservoir and infusion set according to local regulations.
Rewinding your pump
WARNING: Make sure the infusion set is disconnected from your
body before you rewind your pump or fill the infusion set tubing.
Never insert the reservoir into the pump while the tubing is
connected to your body. Doing so could result in an accidental
infusion of insulin, which may cause low BG.
When you rewind your pump, the piston in the reservoir compartment returns to
its starting position and allows a new reservoir to be placed into the pump.
94 Chapter 5