User's Manual
WARNING: Make sure the infusion set is disconnected from your
body before you rewind your pump or fill the infusion set tubing.
Never insert the reservoir into the pump while the tubing is
connected to your body. Doing so could result in an accidental
infusion of insulin, which may cause low BG.
WARNING: Always check your tubing for air bubbles. Continue to
press Fill until the bubbles have been removed from the tubing. Air
bubbles may result in inaccurate insulin delivery.
To fill the tubing:
1. After you load your reservoir and select Next from the Load Reservoir screen,
the Fill Tubing screen appears.
2. Select and hold Fill. Your pump beeps six times to let you know it is
positioning the reservoir. Continue holding Fill until insulin droplets form on
the tip of the infusion set needle, then release. Your pump beeps as it fills the
tubing, and the amount of insulin you are using appears on the screen.
If you get the Max Fill Reached alarm, it means you have used more than
30 units of insulin to fill your tubing. For details, go to Pump alarms, alerts, and
messages, on page 202, and see the description for Max Fill Reached.
3. Select Next to continue.
4. Follow the instructions in the next section to insert the infusion set into your
body before filling the cannula.