User's Manual

3. Follow the instructions in the next section to fill your reservoir.
Filling the reservoir
WARNING: Do not use the reservoir or infusion set if any liquid
gets on the top of the reservoir or inside the tubing connector (as
shown in the image). Liquid can temporarily block the vents. This
may result in the delivery of too little or too much insulin, which
can cause hypoglycemia or hyperglycemia. If any liquid gets on the
top of the reservoir or inside the tubing connector, start over with a
new reservoir and infusion set.
Tubing
connector
Top of
reservoir
WARNING: Always allow your insulin to reach room temperature
before use. Cold insulin can cause air bubbles in the reservoir and
tubing which may result in inaccurate insulin delivery.
96