3. If this is the first time you have inserted a battery in your pump, the Startup
Wizard begins. For more information about the Startup Wizard, see Entering
your startup settings, on page 24. If this is not the first time you have inserted
a battery into your pump, the Home screen appears and the pump resumes
your basal delivery.
Removing the battery
Caution: Do not remove the battery unless you are inserting a new
one, or if you are storing your pump. Your pump cannot deliver
insulin while the battery is removed. After removing an old battery, be
sure to replace it with a new battery within 10 minutes to clear the
Insert battery alarm and avoid a Power loss alarm. If power loss
occurs, you must re-enter your time and date settings.
To remove the battery:
1. Before removing a battery from your pump, clear any active alarms or alerts.
2. Use the belt clip to loosen and remove the battery cap.
Note: Use your belt clip to remove and retighten the battery cap. If
the belt clip is unavailable, you may use a coin.
3. Remove the battery.
4. Dispose of old batteries according to local disposal requirements.
5. After removing your battery, wait until the Insert Battery screen appears
before inserting a new battery.
first steps
First steps 23
MP6025958-022 / A
RELEASED