User's Manual

Do not use the MiniLink™ transmitter (MMT-7703) with the MiniMed 640G insulin
pump as this device does not communicate with this insulin pump.
Do not modify this product, as modification could result in a safety hazard.
Do not use your pump if the screen appears broken or unreadable. In some
instances, impact to the pump can damage the screen while the buttons continue
to function. If the screen is broken or unreadable, do not press any buttons.
Remove the pump and begin using your backup insulin plan per the direction of
your healthcare professional. If the pump is accidentally programmed while the
screen is broken or unreadable, this could result in high or low blood glucose
levels. If your screen is damaged, contact the 24 Hour HelpLine or your local
representative to arrange for shipment of a replacement pump.
Exposure to magnetic fields and radiation
WARNING: Keep the pump away from magnetic fields and avoid
direct contact with any magnets.
If you are going to have an X-ray, MRI, diathermy treatment, CT scan, or other type
of exposure to radiation, take off your pump, sensor, transmitter, and meter before
entering a room containing any equipment of this kind. The magnetic fields and
radiation in the immediate vicinity of this equipment can make your devices
nonfunctional or damage the part of the pump that regulates insulin delivery,
possibly resulting in over delivery and severe hypoglycemia.
Do not expose your pump to a permanent magnet, such as pump cases that have
a magnetic clasp. Exposure to a permanent magnet may interfere with the motor
inside the pump.
Reservoir and infusion sets
Only use reservoir and infusion sets manufactured by Medtronic Diabetes. The
pump has undergone extensive testing to confirm appropriate operation when
used with compatible reservoirs and infusion sets manufactured or distributed by
Medtronic Diabetes. We cannot guarantee appropriate operation if the pump is
used with reservoirs or infusion sets offered by third parties. We are not
responsible for any injury or malfunctioning of the pump that may occur in
association with such use.
■
before you begin
Before you begin 7