User's Manual
4. Review your calculated bolus amount in the Bolus field. If you want to
change the bolus amount, select Bolus and make your desired change.
Remember, if there is a correction bolus amount calculated, you are not able
to give a Square Wave bolus.
Note: If you change your bolus amount, the word “Modified”
appears next to the new bolus amount.
5. Select Next to review your bolus information.
6. Select Square.
The Bolus Wizard screen appears with your bolus amounts.
7. To change the time period over which your bolus is delivered, select
Duration to adjust the time. The duration can be from 30 minutes to 8 hours,
in 15 minute increments.
8. Select Deliver Bolus to start your bolus.
During a Square Wave bolus delivery, the Bolus button on your Home screen
appears as Bolus (S). You can select Bolus (S) to stop the bolus, to see details
on the insulin that has been delivered, or to access the Bolus menu. The
Bolus menu provides access to the Bolus Wizard, Manual Bolus, Preset Bolus,
and Insulin Settings options.
Delivering a Square Wave bolus using Manual Bolus
The Square Wave option is available in the Manual Bolus screen only after you turn
on the Square Wave feature.
To deliver a Square Wave bolus manually:
1. Go to the Manual Bolus screen.
Home screen > Bolus > Manual Bolus
2. Set your bolus delivery amount (in units), and then select Next.
3. Select Square.
4. Select Duration and set the amount of time over which you want your
Square Wave bolus to be delivered. The duration can be from 30 minutes to
8 hours, and is set in 15-minute increments.
78 Chapter 4
MP6025958-022 / A
RELEASED