User's Manual

Stopping and resuming your insulin delivery
Use Suspend Delivery if you need to stop all active basal and bolus insulin
deliveries. When you suspend your insulin delivery, your pump beeps and vibrates
periodically to remind you that insulin is not being delivered.
When you are ready to continue your basal insulin delivery, use the Resume
feature. When you use the Resume feature, your pump starts your programmed
basal pattern, but does not start any previously programmed bolus deliveries.
Note: If you want to stop a bolus delivery only, without stopping your
basal delivery, see Stopping a bolus delivery, on page 88.
WARNING: A bolus delivery or fill cannula that was suspended
does not restart when you resume. To avoid a high BG and
ketoacidosis, always check the pump's Daily History after you
resume insulin delivery to determine the amount that was
delivered. If needed, program a new bolus or fill cannula delivery.
Caution: When using Audio or Vibrate, it is possible that an Audio or
Vibration notification may not occur as expected if the speaker or
vibrator in your pump malfunctions. This is most important when
using the Easy Bolus feature or when your pump is in Manual Suspend
where a missed notification may cause an over or under delivery of
insulin.
Contact the 24 Hour HelpLine or your local representative with any
concerns.
To suspend all insulin delivery:
1. Go to the Suspend Delivery screen.
Menu > Suspend Delivery
2. A confirmation message appears. To suspend your pump and stop all insulin
delivery, select Yes.
The Home screen indicates that your insulin is suspended. Your pump
functions are limited until you resume your insulin delivery.
54 Chapter 3
MP6025958-022 / A
RELEASED