7. SPECIFICATIONS
Power AC 120V, 60Hz
Current 1.5A
Medication Capacity 5ml
Particle Size 0.5 to 10 μm
MMAD 4 μm
Sound Level Around 60 dBA
Maximum Medication capacity 13ml (cc)
Average Nebulization Rate 0.2ml/min
Compressor Pressure Range 25 to 42 Psi (0.17 to 0.29 Mpa)
Operating Pressure Range 8 to 13 Psi (0.056 to 0.090 Mpa)
Liter Flow Range 5~8lpm
Operating Temperature Range 10˚Cto40˚C(50˚F to 104˚F)
Operating Humidity Range 10 to 95% RH
Storage Temperature Range -20˚Cto60˚C (-4˚Fto140˚F)
Storage Humidity Range 10 to 95 % RH
Dimension (L x W x H) 237x120x162mm
Weight 1.70kg
Standard Accessories
Nebulizer, Air Tube, Filters (5pcs), Mouth
Piece, Child Mask.
Protection against electric shock
- Class II Equipment
- Type BF applied parts:
- Mask, baby pacifier
Protection against harmful ingress of water
- IPX0 (Ordinary)
Degree of safety in the presence of flammable anesthetics or oxygen
- No AP/APG (Not suitable for use in the presence of flammable anesthetics or
oxygen)
Mode of operation 20 minutes on 40 minutes off.