Product Manual
7
7. SPECIFICATIONS
Power AC 120V, 60Hz / 230v,50Hz
Current Max. 1.5A / 0.8A
Medication Capacity 5ml
Particle Size 0.5 to 10 μ m
MMAD 4 μ m
Sound Level
≦ 62 dBA (1 m away)
Maximum Medication capacity 13ml (cc)
Average Nebulization Rate
≧ 0.18ml/min
Compressor Pressure Range 17.4 to 29 Psi (0.12 to 0.2 Mpa)
Operating Pressure Range 7.25 to 10 Psi (0.05 to 10.07 Mpa)
Liter Flow Range 5 ~ 8 lpm
Operating Temperature Range 10˚C to 40˚C (50˚F to 104˚F)
Operating Humidity Range 10 to 95% RH
Storage Temperature Range -20˚C to 60˚C (-4˚F to 140˚F)
Storage Humidity Range 10 to 95 % RH
Dimension (L x W x H) 170*136*90mm
Weight 1.18(kg)
Standard Accessories nebulizer, air tube, filters (5pcs), mouthpiece
Protection against electric shock
- Class II Equipment
- Type BF applied parts:
- Mask, baby pacifier
Protection against harmful ingress of water
- IPX0 (Ordinary)
Degree of safety in the presence of flammable anesthetics or oxygen
- No AP/APG (Not suitable for use in the presence of flammable anesthetics or
oxygen)
Mode of operation – 20 min. on & 40 min. off