User manual

34
BlueBP-05
Radiated RF
IEC 61000-4-3
3V/m
80MHz-2,5GHz
3V/m
Recommended Separation distance:
d=[3,5/V1]√P
d=[3,5/3V/m]√P (80MHz –
800MHz)
d=[7/3V/m]√P (800MHz –
2,5GHz)
where:
P is the highest radiated power
disclosed by the manufacturer of
transmitter [W] and d is the
recommended separation distance
[m].
You may observe disturbance
nearby any of those equipment
which has the following indication:
f) Note: in case of frequency 80MHz or 800 Mhz, the formula for the higher range is
applicanle.
g) Note: these are guidelines. Actual conditions may vary.
Recommended separation distance
Meditech ambulatory blood pressure monitors are intended to be used in electromagnetic
environment with controlled RF disturbances. The purchaser or user of the devices may help
to reduce electromagnetic disturbances by defining the separation distance between the
transportable or mobile RF telecommunication equipment (transmitters) and the device,
depending on the highest output power of the telecommunication equipment.
Separation distance in function of the frequency of the transmitter (m)
The highest output power
of the transmitter (W)
150KHz-80MHz
d=(3,5/V1)√P
80MHz-800MHz
d=(3,5/E1)√P
800MHz-2,5GHz
d=(7/E1)√P
0,01
Not applicable
0,12
0,23
0,1
Not applicable
0,38
0,73
1
Not applicable
1,2
2,3
10
Not applicable
3,8
7,3
100
Not applicable
12
23
If this table does not contain the highest output power of the transmitter, the d separation
distance [m] can be calculated by the formula, depending on the frequency of the transmitter,
where P is the rated highest output power of the transmitter [W].
1. Note: in case of frequency 80MHz or 800 MHz, the formula for the higher range is
applicable.
2. Note: These are guidelines. Actual conditions may vary.