User manual
33
ABPM-05
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
3A/m
3A/m
Power frequency
magnetic fields are at
levels characteristic
of a typical location
in a typical
commercial and/or
hospital
environment.
ABPM-04
BlueBP-05
Note: U
T
is the nominal voltage of mains.
Monitors
Immunity
test
IEC60601 test
level
Compliance
level
Electromagnetic
environment
ABPM-05
Conducted RF
IEC 6100-4-6
3V eff
150KHz-80MHz
Not applicable
Portable and mobile RF
communications equipment are
used no closer to any part of
Meditech ABPM-05/ABPM-04,
including cables, than the
Recommended Separation
Distance calculated the formula
written below.
Recommended Separation
distance:
d=[3,5/V1]√P
ABPM-04
BlueBP-05
3V
eff
Portable and mobile RF
communications equipment are
used no closer to any part of
Meditech BlueBP-05, including
cables, than the Recommended
Separation Distance calculated
the formula written below.
ABPM-05
Radiated RF
IEC 61000-4-
3
3V/m
80MHz-2,5GHz
3V/m
d=[3,5/3V/m]√P;
(80MHz – 800MHz)
d=[7/3V/m]√P;
(800MHz – 2,5GHz)
where:
P is the highest radiated power
disclosed by the manufacturer of
transmitter [W];
d is the recommended
separation distance [m].
ABPM-04