15. EMC information
Medical electrical equipment should be used with precautions according to EMC, and must
be installed according to the EMC notices disclosed in this manual as mobile RF transceivers
could adversely affect it.
Electromagnetic emission
Meditech ambulatory blood pressure monitors are suitable for use in the specified
electromagnetic environment. The purchaser or user of the device should assure that they are
used in an electromagnetic environment as described below.
Monitor
Emission test
Compliance
Electromagnetic
environment
ABPM-
05
Radiated and conducted RF
emission CISPR
11
Group 1
ABPM-05/ABPM-04/BlueBP-
05 uses RF energy only for their
internal function. Therefore, the
emission is very low and it is not
likely to cause any interference in
neraby electronic equipment.
Except for Bluetooth
communication usage.
3
ABPM-
04
BlueBP-
05
ABPM-
05
Radiated and conducted RF
emission CISPR 11
Class B
ABPM-05/ABPM-04/BlueBP-
05 is suitable for use in domestic
establishments and in
establishments directly
connected to the low voltage
power supply network which
supplies buildings used for
domestic use.
ABPM-
04
BlueBP-
05
ABPM-
05
Harmonic emission
IEC61000-3-2
Not applicable
---
ABPM-
04
BlueBP-
05
ABPM-
05
Voltage fluctuations/ Flickers
IEC61000-3-3
Complies
ABPM-05/ABPM-04 is suitable
for use in establishments directly
connected to a public low
voltage mains network.
ABPM-
04
BlueBP-
05
Not applicable
---
3
The Bluetooth radiofrequency communication is carried out at the 2.4GHz frequency.
(class2, Bluetooth 2.0+EDR)