User manual

16
8. Apply the cuff to the patient with the device placed in the pouch.
9. Give the patient diary to the patient along with detailed instructions about the rules and
the use of the device.
Manual programming (ABPM-05 only)
ABPM-05 can be programmed without a PC by using the device buttons.
Programming options
Plans are stored in the inbuilt memory of the device and they cannot be changed. The
following three measurement plans can be selected during the programming of the device:
- First plan: measurements every 15 minutes at day and every 30 minutes at night.
- Second plan: periods with 20 minutes at day and 40 minutes at night.
- Third plan: 30-minute periods indepentent of day or night time.
Other settings are the same in all the three plans: undecided cuff size, 300 mmHg pressure
limit, LCD display enabled, manual day/night shift disabled. Daytime starts at 6:00, while
nighttime starts at 22:00, special session is disabled. Patient data can be selected or created
later in CardioVisions database.
Measurement timing
The first measurement has a controlling purpose and it starts in the second minute after
programming, then other measurements cannot be started in the next five minutes. The rest
of the measurements are taken at specific 15/20/30/40 minute intervals and there are
measurements at the 6:00 and 22:00 hour shifts. The last measurement is exactly 24 hours
after the second measurement.
Manual programming step by step
- Press and hold START and DAY/NIGHT button at the same time continuously.
Measurement frequency of the measurement plans will be displayed after 10 seconds for 3-3
seconds.
- In order to choose the measurement plan, you should release the buttons while the specific
plan is displayed. You will hear two beeps and the LCD will display four blinking „o” letters,
which indicates that programming the device is in progress. After a successful programming
you can hear 5 beeps, and the selected plan can be seen again. If programming fails for some
reason, the E90 error code will be displayed on the LCD.
In case of manual programming there is no time setting. If the time setting is imprecise, the
time of measurements may be false. If you want to use the manual programming function,
do not leave the device without batteries for a longer period. If it happens so, insert batteries
again into the device, set the inner clock by programming the device by a PC and leave the
batteries in the device.
Software
The devices programmed manually can be used by CardioVisions software version 1.13 or
later versions.
LCD displays for manual programming
Normal status: time is
displayed.
2. measurement plan:
20/40 minute
day/night intervals