User manual

11
6. Technical parameters
Technical parameters
ABPM-05
BlueBP-05
ABPM-04
Power supply
2 AA rechargeable NiCd or NiMH batteries or 2 AA alkaine
batteries
4 AA rechargeable Nicd or
NiMH batteries or 4 AA
alkaine batteries
Display
liquid-crystal
Data storage
internal solid state memory
Data transmission
serial or USB cable
serial Bluetooth transfer
serial or USB cable
Operating environment
temperature
10-45 °C
humidity (non condensing)
10-95%
10-80%
10-95%
atmospheric pressure
70-106 kPa
83-103 kPa
Storage conditions
temeprature
-20 - 50 °C
humidity (non condensing)
10-95%
10-80%
10-95%
Size
70*99*30 mm
82*124*33,5 mm
Weight (batteries included)
240g
330g
Blood pressure measurement method
oscillometric
Blood pressure maximum stirage
over 600 measurements
Measurement range
blood pressure: 30-260 mmHg (4-35 kPa)
pulse: 40-200 beat/minute
Passive accuracy
+/- 3mmHg (0,4 kPa) or +/- 2% of measured value (stability: 2 years)
Blood pressure measurement accuracy
measuring algorithm validated to BHS and AAMI protocol
Pressure senzor
stepwise deflation
Inflation
automatically controlled pump
Safety
maximum inflation 300 mmHg (40 kPa), independent safety release valve
Deflation and rapid air release
automatic pressure release value