User's Manual

Page7 / 35
Storage and Handling of the System
Storage and Handling of the Control Solution
Important Information for Caring the Meter
Clean your Medisign MediKeto Plus 82 BT blood glucose and β-ketone meter once per week at a
minimum.
Do NOT get water inside the meter. NEVER immerse the meter or hold it under running water or spray
any cleaning solution directly onto the meter.
Do NOT try to get moisture on the test strip, test strip slot or serial port.
Do NOT use glass cleaners or household cleaners or corrosive liquid (e.g. Benzene, Acetone) on the
meter.
Please use a soft cotton cloth to wipe the system.
Squeeze the excess liquid from the cloth before you wipe the meter.
The meter must be turned off when you clean the meter.
The control solution should be stored at indoor area between 4 ~ 30℃. Do NOT freeze.
The control solution may be stored in a refrigerator between 4 ~ 8℃. But it must be brought to room
temperature before using.
The control solution is good three (3) months after first opening date or until the expiration date,
whichever comes first.
Record the discard date on the control solution vial and discard remained control solution after that date.
The control solution test is recommended to be done at room temperature (20 ~ 25℃).
Keep this system dry and keep away from the heat and direct sunlight.
Avoid leaving the meter or test strips in extremely hot or cold places, such as near a heat source or in an
extremely hot or cold car.
Do NOT store or use the meter or test strips at high humidity levels, such as in the bathroom or kitchen.
Store the test strips in the original vial between 4 ~ 30℃, at indoor area. Do NOT freeze.
The test strip may be stored in a refrigerator between 4 ~ 8℃. But it must be brought to room
temperature before using.
When properly stored, unopened test strips are stable until the expiration date printed on the vial or the
foil packet.
The Medisign GH82 test strips in the vial are good six (6) months after first opening date or until the
expiration date, whichever comes first.
The Medisign MediKeto 82 test strips in the vial are good three (3) months after first opening date or
until the expiration date, whichever comes first.
Record the discard date on the test strip vial and discard any unused test strip after that date.
The meter and the test strips are designed to be used within a temperature range of 10 ~ 40℃. Please
keep the system operating temperature at a range between 10 ~ 40℃ for more than twenty (20) minutes
before testing.
Do NOT test if there is condensation (water build-up) on your meter and the test strip vial or the foil
packet. Move your meter or a test strip vial or a foil packet to a cool, dry spot. Wait for the meter and a
test strip vial or a foil packet surface to dry before testing.
The meter, test strips, and control solution should be handled at the same temperature.
Do NOT handle the test strips with wet or dirty hands.
For accurate results, make sure your hands are clean and dry before retrieving test strip from its foil
packet.
Do NOT allow dirt or dust in the test strip slot in order to avoid affecting the system’s performance.
Dispose of used test strips, lancets according to instructions from your healthcare professional.
Dispose of used batteries according to local regulations.
Do NOT destroy, repair or transform the system at any time. If the equipment is used in a manner NOT
specified by the manufacturer, the protection provided may be impaired.
Handle the meter with care severe shock, such as dropping the meter, could damage the electronics.
The Applicable Pollution Degree of the intended environment for this system is Pollution Degree 2.