User's Manual
Page33 / 35
█ Troubleshooting
What is happening
What you should do
The meter does NOT work after the
blood is applied.
Make sure that the test strip is insert into the test strip slot
completely.
See if the test strip is covered in debris or lint.
Make sure the blood is applied to the test strip correctly.
Make sure the test strip is inserted correctly and repeat the test.
The test result is abnormal.
See if the test strip is covered in debris or lint.
Check if the test strip is within the validity period.
Make sure the test strip has NOT been stored out of its vial.
Make sure the test strip has NOT been used before.
Make sure the code on the meter display matches with the test
strip code.
The thermometer symbol appears
on the screen.
Put the meter and test strips in a location within the operating
temperature range (10 ~ 40℃) for at least twenty (20) minutes
and repeat the test. Do NOT raise or lower the meter
temperature artificially.
If the thermometer symbol is constantly showing, contact
Customer Service.
The meter does NOT work after the
test strip is inserted.
Make sure the test strip is completely inserted into the test strip
port.
Make sure the batteries are inserted correctly, with the + sign
facing upwards.
14. System Specifications
Measuring Range (Glucose)
20 ~ 600 mg/dL
Measuring Range (β-ketone)
0.1 ~ 8.0 mmol/L
Measuring Time (Glucose)
6 seconds
Measuring Time (β-ketone)
8 seconds
Memory capacity
500 test results (include date and time)
Sample Type (Glucose)
Fresh capillary and venous whole blood
Sample Type (β-ketone)
Fresh capillary(only fingertip) and venous whole blood
Operating condition
10 ~ 40℃
Sample volume
0.5 ㎕
Screen type
Backlight LCD
Dimensions (mm)
90 (L) x 53 (W) x 18 (H)
Weight
Approximately 52g (with batteries)
Power supply
DC 3V (CR2032) Lithium Battery
Data Output
Bluetooth, USB Data Transfer Cable
Wireless Frequency
2.4 GHz Band