Troubleshooting
The data can be deleted from the memory selected by pressing and holding the
SET button 4
for about 7 seconds. After 7 seconds, the display
3 shows
CL” which indicates that all data have been deleted.
The unit switches off automatically after about 3 minutes, if no button is
pressed during this period. It can also be switched off by pressing the START
button 2 . The time is on permanent display. A short press of the
MODE
button 5
will bring up the date briefly in the display, then return to the time.
ERROR Error during measurement
Determining pulse
Batteries low. Please replace batteries.
If the display shows one of the following error messages, this means that the
unit was unable to accurately determine blood pressure.
Error messages
Possible causes and solutions
ERR 1
Measurement of the systolic blood pressure was completed.
Either the connection with the hose has been interrupted or no pulse could be
determined.
Check the connections between the cuff and the main unit.
ERR 2
Non-physiological pressure has been detected.
Cause: the arm was moved during the measurement.
Repeat the measurement, keeping your arm still.
ERR 3
If the cuff takes too long to inflate, it is possible that the cuff is incorrectly
positioned or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
ERR 4
The difference between systolic and diastolic pressure is abnormally high.
Repeat the measurement under relaxed conditions, following the instructions
exactly.
If the same error occurs repeatedly with further readings, please contact
MEDISANA.
ERR 5
An average value could not be calculated due to unstable conditions while the
readings were being taken.
Repeat the measuring process, keeping your arm still and not speaking.