Contents Preparations nets beers senses serene 3 Before use Before use Names of components Charging the battery. Power indicator Assembling Using the setup mode “ A. Digital Scopes (DOC 300S/ DOC 1008), Digital speculate (DGC 100}, Digital Anterior Scope (DEA 100 and DEA 100 with ILS 100), Endoscope adapter (Adapter 300) ..31 B. DEC 100/ EEC 100 (Digital Eye Funds Camera) C. DEA 200P (Digital Anterior Scope) D.
Preparations Before use Prior to installation and start-up of the Horus Scope, carefully read the user manual. As with all technical devices, the proper function and safety operation of this device depend on the user complying with the safety recommendations presented in these operating instructions. in addition, please make sure it does not appear damaged or broken. If there are breaks on the outer casing or other visual defects, please contact the manufacturer or a certified service facility.
Safe eye screening £1 While no acute optical radiation hazards have been identified with the camera, it is recommended that the intensity of light directed into the patient's eye be limited to the minimum level necessary for diagnosis. infants, shakes, and persons with diseased eyes are at greater risk. The risk may also be increased if the person being examined has had any exposure to the same instrument or any other ophthalmic instrument that uses a visible light source within the previous 24 hours.
Bb bb All cables can only be used by the manufacturer or distributors. and that are likely to affect compliance of the ME EQUIPMENT or ME SYSTEM with the requirements of IEC/ EN 60601-1-2 Clause 7 (EMISSIONS) and Clause 8 (IMMUNITY). WARNING: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
DGC 100 at operation. Charging the battery po Bb Ps The time required for charging varies depending on the conditions of battery usage. Charging takes longer at high or low temperatures and when the battery has not been used for some time. The battery will get warm during charging and for some time thereafter. The battery will be drained completely if not used for long periods of me, even after being charged. Only use Li-ion Battery 3.
Make sure the jig is completely locked by pushing it downward. & The slit lamp jig is only suitable for DEC 100 and EEC 100. Protection & Do not attempt to remove the cover from the product to prevent the product from malfunctioning. Pits No modification of this device is allowed. The performance would be subject to any modification and may cause hazardous radiation exposure.
pate It is recommended to clean the funds lens with a cleaning cloth or lens cleaning tissue, such as THOR LABS Inc. (www. blackthorn) lens cleaning tissue. if a replacement for the eye cup or contact plate is needed, please contact the manufacturer or retailer. Clean the aye cup or contact plate before each use: 2 Disinfect the eye cup or contact plate with soft cloth moistened with alcohol (75% ethyl alcohol). NOTE The device Is not intended to be sterilized.
Operating Environment = Ambient temperature: 10°Cto +35°C ¢ Relative humidity: Atmospheric pressure: 800hPa to 1013hPa * Shock (without packing): 10 G, duration 6 ms Environment for Storage = Ambient temperature: ~10°C to 455°C & Relative humidity range: Atmospheric pressure: 700 hPa to 1013 hPa NOTE It is recommended to remove the battery if the device is stored over two weeks.
Names of components Scope of Delivery Control unit Mi Is Horus Scope DSC 300P 1. Battery 2. Power adapter 3. Mini USB cable 4. Micro HDMI cable 5. Charging station 6. Micro SD card 7. Portable chin rest (Optional) 8. Image Management System, Model: SA 1 standalone (option) 9. Image Management System, Model: SB 1 {free download) Digital Scopes Mi Is Horus Scope DOC 300S Mi Is Horus Scope DOC 100S 1.
Product name Model Name Control unit Mi Is Horus Scope DSC 300P An interchangeable lens digital camera used to record digital photographs and video of human eye, human ears, human skin and human body. Digital Scopes Mi Is Horus Scope DOC 300S Mi Is Horus Scope DOC 1008 A digital hand-held scope used to record digital photographs and video of the human ear's canal and tympanic membrane.
User interface Control Unit {Mi Is Horus Scope DSC 300P) Top view » Front view » 12 Power button Lens locker 3.
Bottom view » —— Mini USB Rear view » B Pogo pad Micro HDMI Strap hook Lens assembling mark Cover glass and sensor Lens connecting pins Focus wheel Battery cover Repair hole SD slot 13
introduction of scopes MIS Horus Scope DOC 3008/ DOC 1008 [Digital Scopes) DOC 3008 Structure Spiel oe Fatso very Technical Description Focus Range 7~20 mm (Typical) Dimension 4.3 X4.
MIS Horus Scope DEC 100/ EEC 100 {Digital Eye Funds Camera) DEC 100 Si i, Signal Pa Structure Cm 4 frat ad Lens wizard Contact Sing Contact Surface Technical Description View Angle 40 Degree (Typical) Dipterous -20 ~ +20D (Typical) Dimension 4.3 X4.3X12.
MIS Horus Scope DEA 100 [Digital Anterior Scope) Structure Lens Cover LED Switching Button Signal Pad Contact King tens mark Position Bar Technical Description 11 mm mm mm (Diagonal) @ View Angle Working Distance 30 mm (Typical) Dimension 4.3 X4.3X 4.5 cm (Typical) Weight 91 Grams (Typical) Camera / video light source Natural White/Blue Light Emitting Diode {LED} Switch to blue LED by the LED Switching Button to LED Switching Button take fluorescence image of cornea.