User's Manual

19
Specifications
Physical
Size: 4.6 x 2.0 x 1.3 inches (HWD)
Weight: 3.7 oz less batteries, 5.0 oz with batteries
RF Characteristics
Transmission type: Digital
Frequency range: 608 – 614 MHz WMTS band
Frequency tolerance: Plus or minus 5 ppm
Output power: 5dbm
ECG
Leads transmitted: Lead II, V
Number of electrodes: 3 or 5
Leads transmitted simultaneously: 2
Input impedance (megohms): >20
Frequency response (Hz): 0.05 – 100
Pacer ID: Yes, 2mV to 700mV, 100usec. To 2msec.
Dynamic range: Plus or minus 5mV
DC Offset: Plus or minus 400mV
Common Mode Rejection Ratio: 80db @ 60Hz minimum
Defibrillation Recovery Time: within 8 seconds
Power
Battery type: 3 1.5V AAA, Alkaline
Battery life (days): Minimum 2.5 days with alkaline battery est.
Reverse battery protection: Yes
Low battery signal: Yes
Environmental
Temperature range: 10 – 40 degrees Celsius
Humidity range: 15% - 95% relative humidity
NOTE: No maintenance or calibration required. If returned to MDE, the Angel
MP must be shipped in packaging specified by MDE.