User's Manual
16
5. Tape each lead wire to skin to help maintain stable ECG baseline.
6. Insure electrodes are placed in proper locations.
7. Confirm appropriate alarm status and limits exist for patient based on clinical assessment.
Power Down
Discontinuation of the Angel MP transmitter requires that all leads and cables be removed from
patient.
Angel MP's power remains on while batteries are in battery compartment and knob is in the ON
position. Turn knob to OFF and remove batteries from Angel MP for long-term storage or non-
use.
Receiver must be TURNED OFF at the Central Station to halt lead alarms and display. This is
done at the Central by touching patient’s window. An ECG window display appears on lower
screen. Touch Transmitter tab. Turn receiver to OFF.
Caring for the Angel MP
• For cleaning, use only a lint-free, nonabrasive cloth which has been slightly dampened with
mild detergent, a glass cleaner, or a non-alcoholic based solvent.
• Avoid harsh cleaning solutions such as isopropyl alcohol or other solvents that might harm
plastic surfaces. Avoid anti-bacterial cleaning solutions, especially those containing Phenol.
• Do not immerse unit or any components in liquid.
• Do not spray liquids directly on the Angel MP.
• Do not allow liquids to penetrate connectors or the Angel MP's case.