User Manual

7
Getting Started
About Your Vision Central Station
The ESCORT Vision Central Station provides communication, centralized surveillance and documentation
of patient vital signs and arrhythmia monitoring for up to 16 patients.
CAUTION:
Pressing ON/OFF switch or System Reset button will cause a
complete loss of data in the Random Access Memory (RAM) of Vision
Base. Stored patient history will be lost.
Turning the Vision On
1.
Press Vision Base power switch to activate base.
2.
Press Vision Display power switch to activate display.
3.
Toggle power switch on Link Auxiliary Base to the ‘
I
’ position to activate.
4.
Adjust Vision Display brightness and contrast controls for comfortable viewing.
Turning the Vision Off
You will usually not need to turn Vision off. However, under special circumstances such as scheduled
power outages, field service, or when the need to move equipment arises, you may have to turn Vision off.
MDE recommends only system maintenance personnel shut system down. However, if you need to turn
system off, be sure to follow prescribed power-down sequence.
TO POWER SYSTEM DOWN:
1.
Touch
SYS SETUP
.
2.
Touch
SYSTEM OPTIONS
.
3.
Touch
VISION POWER DOWN
.
4.
Touch
YES
to proceed. When you see message
It is now safe to turn off your computer
turn front
panel power switch off.