User Manual
26
Alarm, Save, Print, or Record Anomalies
Problem May Be Caused By Try This...
Alarm does not sound;
central station does not
save, or does not record.
Arrhythmia has not been
turned “on” for patient.
Verify setup for patient.
Ensure arrhythmia
monitoring option is “on”
and limits are set
appropriately.
Event limits are set to
inappropriate values.
Verify setup for patient.
Ensure proper settings are
“on” and limits are set
appropriately.
Desired event settings are
not turned “ON” on
Alarm/Arr popup.
Verify setup for patient.
Ensure proper settings are
“ON” and limits are set
appropriately.
All beats in event did not
meet algorithm’s criteria
for classifying the event.
Use Full Disclosure option
to verify classification of
beats in arrhythmia event.
Print Buttons do not appear
on popup.
“Have Printer” is not
selected on Configuration
screen.
Access Configuration
screen (this requires a
password). Check the
“Have Printer Box.”
Miscellaneous Conditions
Problem May Be Caused By Try This...
No Start/Stop buttons
appear on NBP popup.
“Allow Start/Stop on NBP”
may not be selected on
Configuration screen.
Access Configuration
screen (requires password)
and touch
Allow
Start/Stop on NBP
to put a
check in box.
Patient ID does not show in
patient header when
admitting a patient.
“Pat-ID in Patient Header”
may be unchecked on
Configuration screen.
Access Configuration
screen (requires password)
and touch
Pat-ID In
Patient Header
to put a
check in box.