User Manual
24
Arrhythmia Anomalies
Arrhythmia anomalies can occur for a variety of reasons. Often they are traced to problems with patient
electrodes, lead wires, or cables. Always ensure patient ECG signal is as clean as possible.
Problem May Be Caused By Try This...
Anomalous event
classifications.
Noisy ECG signal causing
false event detection and
classification due to
attachment, motion, or
electrical artifact.
Ensure ECG signal is clean
by using good setup and
lead placement techniques.
If ECG is still noisy, check
electrodes, lead wires, and
patient cable for possible
failure.
Minimize any possible
muscle motion, contact, or
electrical artifact.
Lead change or change in
patient’s rhythm has altered
learned dominant beat.
Relearn current dominant
rhythm at central station.
T-waves and/or P-waves
are being detected and
classified.
Select a lead that
maxi mizes R-wave and
results in P-waves and T-
waves no larger than 3 mm
on recorder. Relearn ECG
after making adjustments
via central station.
Atrial fibrillation or flutter
are difficult rhythms. The
frequent rate changes and
distortion of QRS may
result in miscalculation of
beats and events.
Adjust limits to minimize
false Run, Missed Beat, and
SVTach classifications.