User Manual
10
Set Arrhythmia and Vital Sign Alarms
1.
Touch
ARR/ALARM SETUP
.
2.
Touch appropriate patient display tile.
3.
Default tab is
ARRHYTHMIA
settings. Touch
BEDSIDE
tab to access vital sign alarms,
TELEMETRY
tab to access transmitter settings, and
ST
tab to access ST settings.
4.
Touch alarm related setting (
Alarm, Level, Record, Save, Value
) you want to change for specific
alarm events.
5.
Touch
ARRHYTHMIA ON/OFF
to turn arrhythmia processing
ON
or
OFF
for that patient.
6.
Touch
RELEARN
to cause arrhythmia algorithm to relearn a patient’s normal rhythm.
7.
Touch
STANDBY
to suspend arrhythmia processing for a specific patient for 3 minutes.
8.
Touch
BEDSIDE ALARM TONES
to turn bedside monitor alarm tones
ON
or
OFF
.
NOTE:
This setting does NOT affect central station alarm tones.
Adjust BEDSIDE Vital Sign Settings
1.
Touch numeric display tile of ANY vital sign to be adjusted (for example, HR, NBP, RSP. etc.).
2.
Adjust vital sign functions as desired. Access to common functions include:
•
Touch
HR
numeric tile for desired patient, then touch
ECG
tab to change ECG lead
configuration, turn pacer detection ON or OFF, and adjust size of ECG waveform.
•
Touch
NBP
numeric tile for desired patient to access ability to START/STOP a cuff measurement
and adjust desired automatic cuff measurement interval.
•
Touch
RSP
numeric tile for desired patient to adjust size of Respiration waveform.