User's Manual
A-1
APPENDIX A
SPECIFICATIONS
GENERAL
Patient Safety
Designed to meet the requirements of CSA, UL 2601 and IEC 601-1. Transmitter complies with
radio frequency exposure limits specified in FCC sections 1.1307 and 2.1093.
Water resistant.
Defibrillator protection up to 5 KV.
Power Requirements
Operating Voltage 3 V.
Battery Type Two (2) 1.5V AA batteries, alkaline.
Battery Operation Time 36 hours with continuous SpO2, minimum.
72 hours with ECG only, minimum.
Environment
Operating Temperature +10° C to 40° C
Storage Temperature -40° C to 70° C
Operating Relative Humidity 30% to 75%
Storage Relative Humidity 10% to 100%
Dimensions
Height 13.2 cm (5.2 inches).
Width 7.4 cm (2.9 inches).
Depth 2.54 cm (1.0 inch).
Weight 127.5 g (4.5 ounces) without batteries.
Remote Communication
FCC Certification EHC20701
Rated RF Output Power 5 dBm
Frequency Range WMTS transmit band (608-614mhz).
Number of Channels 12.5 kHz frequency resolution with the ability to operate at
75 kHz channel separations.
Display
Type LCD monochrome display with backlight.
Screen Size 3.8 x 1.3 cm (1.5 x 0.5 inches).
Sweep Speed 20 mm/sec
Waveform Display Height 7.5 mm