User's Manual

1-4
(1) Available BIO Configuration settings are: Channel ID, Language, LCD
Intensity, SpO2 Avg., SpO2 Mode, Nurse Call Timeout, Remote Record
Enable, 3- or 5-Lead Selection and 3-Lead Default.
(a) SELECT CHANL. This selects the channel ID for connection to the
Vision Central Station or bedside monitor.
(b) LANGUAGE. The only language supported at this time is English.
(c) LCD BIAS. This value will adjust the LCD contrast setting with an
available range of 0 - 31. The default is 15.
(d) NURSE. This selects the duration of the Nurse Call timeout. Once
initiated, the Nurse Call will remain active until manually cleared (by
pressing the Waveform key) or until the selected timeout value has
elapsed. The available timeout values are 1 (minute), 2 (minutes), 3
(minutes), 4 (minutes), 5 (minutes), LATCHING and OFF. The default
is 1 minute.
(e) RECORD. This selection will enable/disable the remote record
function. The default is ON (enabled).
(2) Advance through the available options by pressing the Waveform key.
(3) To change the settings use the Nurse Call key to Increment through the
options or answer YES and the Remote Record key to Decrement through the
options or answer NO.
(4) Once all the configuration settings have been scrolled through, the SAVE/
EXIT query will appear. If no changes were made only the EXIT query
appears. To save the changes answer YES to SAVE, to discard the changes
answer NO.
b. NURSE Mode. To enter the NURSE Mode, place the VISION TelePak in normal
mode of operation with the display screen ON then hold down both the Nurse Call and
Remote Record keys for three (3) seconds. The display will change to the first Nurse
Configuration page.
(1) Available NURSE Configuration settings are: SpO2 Mode, SpO2 Avg., 3- or
5-Lead Selection and 3-Lead Default.
(a) O2 MODE. This selects the O2 interval time. Available selections are
1 minute, 2 minute or OFF (continuous). The Default setting os OFF
(continuous).
(b) O2 AVG. This selects SAT/PR value averaging. The available
selections are SLOW (8 seconds) and NORMAL (4 seconds). The
default is NORMAL
(c) LEAD MODE. This sets lead configuration to 3-lead or 5-lead mode.
The default is 5-lead.
(d) 3LD SELECT. This selects the 3-lead configuration. Only I and II are
available. The default is II.
(2) Advance through the available settings by pressing the Waveform key.
(3) To change the settings use the Nurse Call key to Increment through the
options or answer YES and the Remote Record key to Decrement through the
options or answer NO.
(4) Once all the configuration settings have been scrolled through, the SAVE/
EXIT query will appear. If no changes were made only the EXIT query
appears. To save the changes answer YES to SAVE, to discard the changes
answer NO.