User manual
Quick guide P9-0135-Q
 10 of 20
Version 1.0.0/2012 
Medema Production A/S
Battery indicator
The table below indicates the fault types which the battery indicator can display 
in the case of an electronic fault on the Mini Crosser. Count these and read off in 
the table what kind of fault they indicate. However, it is recommended that you can 
quote the number of the fl ash code when contacting the supplier.
LED 
CODE
PROGRAMMING 
LCD DISPLAY
EXPLANATION POSSIBLE CAUSE
1,1 HARDWARE 
FAILSAFE 2
HARDWARE 
FAILSAFE 4
Loss of power
Overcurrent fault
1. Short circuit in motor or 
 motor cable
2. Failure of control unit
1. Short circuit in motor or 
 motor cable
2. Failure of control unit
1,2 HARDWARE 
FAILSAFE 1
HARDWARE 
FAILSAFE 3
PRECHARGE 
FAULT
HW FAILSAFE
EEPROM fault
Main contactor fault
Precharge fault 
Motor voltage fault
1. Fault in or failure of 
 EEPROM
1. Main contactor welded
2. Fault in driver in main 
 contactor
3. Fault in coil in main 
 contactor
1. Fault in internal control 
 unit
2. Low battery voltage
1. Motor voltage does not 
 fulfi l throttle requirements.
2. M1 or M2 power short-
 circuited to B- or B+.
3. Internal motor short circuit.
4. Failure of control unit
2,1 SRO SRO fault 1. Wrong sequence of KSI, 
 power activation and input 
 data for direction
2. Selection of wrong SRO 
 type
3. Direction switch circuit 
 open