Specifications
8. Symbol Definitions:
0﹉﹉﹉
-Zero point is over the setting range
0﹍﹍﹍
-Zero is below the setting range
Ad﹉﹉﹉
-Analog digital converter chip over max. range
Ad﹍﹍﹍
-Analog digital converter chip below min. range
﹉﹉﹉﹉ -Weight signal is too large
﹍﹍﹍﹍ -Weight signal is too small
EEP.E1
-Config parameters incorrect (no set, no calibration, over normal range, etc.)
EEP.E2
-User parameters are incorrect
CAL-Px -Calibration point
CAL.Er
-Error in calibration
CAP.--
-The setting full capacity will be displayed
Cx.y -No. x configuration parameter is set to y
Ux.y -No. x user parameter is set to y
Lo.bAt
-Battery voltage is below 3.6V
9. Trouble shooting:
SYMPTOM PROBABLE CAUSE REMEDY
Does not turn on.
1. AC adapter is not securely connected
2. Low battery
3. Indicator is damaged
1. Re-plug the AC adapter or rotate the plug
to securely connect to the scale
2. Replace the batteries
3. Replace with a new indicator and preform
calibration
Ad﹉﹉﹉
Ad﹍﹍﹍
1. The cable from platform to indicator is not
correctly connected, or disconnected, or
short circuit
2. Indicator is damaged
3. Load cell cable is broken
4. Load cell is damaged
1. Check the cable or replace with a new
cable
2. Replace with a new indicator and perform
calibration.
3. Return the scale for repair
0﹉﹉﹉
Indication is out of key zero range
Reduce the weight on platform, till the
indication is within the key zero range.
0﹍﹍﹍
Weight reading below Power On Zero limit.
1. Loosen the shipping protection screws
2. Check whether an object is stuck between
scale base, if yes, remove the object.
3. Perform zero calibration.
﹉﹉﹉﹉
1. Weight reading exceeds Overload limit
2. The weight value cannot be displayed in
the current unit of measure because it
exceeds 6 digits.
1. Reduce load on the scale until a weight
value is displayed.
2. Use a more appropriate unit of measure.
﹍﹍﹍﹍ Weight reading below Under load limit
1. loose the shipping protection screws
2. Perform zero calibration
PS‐103 www.measuretek.net .16