Specifications

Para-
meter
Default Setting
12R975
setting
12R976
setting
12R977
setting
12R978
setting
12R979
setting
C8 2
Zero range for ZERO/ON/OFF button:
0 –power-on zero -point ±1%FS;
1 - power-on zero –point ±2%FS;
2 - power-on zero –point ±3%FS;
3 - power-on zero –point ±4%FS;
4 - power-on zero –point ±5%FS;
5 - power-on zero –point ±10%FS;
6 - power-on zero –point ±20%FS;
7 - power-on zero –point ±50%FS;
8 - power-on zero –point ±100%FS;
9 - no limitation;
6 6 6 6 6
C9 0
Select which zero point will be used after scale is powered
on and current weight signal is within the power-on
zero-point range:
0 - Current weight;
1 - Calibration zero point;
2 - Power-off zero-point (power-off tare weight as current
tare weight).
0 0 0 0 0
C10
0
Select which zero point will be used after scale is powered
on and current weight signal is NOT within the power-on
zero-point range:
0 - Prompt power on zero point is over range;
1 - Current weight;
2 - Calibration zero point;
3 - Power-off zero-point (power-off tare weight as current
tare weight).
0 0 0 0 0
C11
6
Zero tracking range:
0=no tracking; 1=±0.25d; 2=±0.5d; 3=±1d; 4=±1.5d;
5=±2d; 6=±3d; 7=±4d; 8=±5d; 9=±6d.
6 6 6 6 6
C12
2
Digital filter intensity:
0= very weak; 1=weak; 2=middle
; 3=strong;
2 2 2 2 2
C13
5
Motion check range:
0=±0.25d, 1=±0.5d2=±1d3=±1.5d4=±2d
5=±3d6=±4d7=±5d8=±6d9=±7d
4 2 4 2 4
C14
1
Max. weight display:
0=FS+0d; 1=FS+9d;2=101%FS; 3=102%FS;
4=105%FS; 5=110%FS; 6=120%FS; 7=150%FS;
8=200%FS; 9= No limitation
1 1 1 1 1
C15
0
0=BMI function is disabled
1=BMI function is enabled
0 0 0 0 0
C16
0
0= Do not recover parameters (Cxx & Uxx) to default set
1= Recover all parameters to default setting
0 0 0 0 0
PS103 www.measuretek.net .9