Specifications

6. Configuration Parameters Setup
1. When the scale is off, press and hold the ZERO/ON/OFF and PRINT/HOLD keys until ‘C0NF’ is shown, which
indicates that the scale is in Configuration parameter setup mode.
2. During setup mode, press the UNIT key to change the flashed digits, and use the HOLD/PRINT key to shift the
flashed position. Press the TARE key to confirm the flashed digits. Press the ZERO/ON/OFF key to exit this
mode.
3. Summary of Configuration Parameters Setting:
Para-
meter
Default Setting
12R977
setting
12R979
setting
C1 3000
Display Resolution for main weight unit: 100–20000
1650 2200
C2 3307
Display Resolution for second weight unit (if main unit is kg or lb, then
second unit is lb or kg): 100–25000
Note: it must be equal or less than 1.25×C1
1500 2000
C3 0
Division select: 0 – 1; 1 2; 2 5;
0 1
C4 0
Decimal point: 0 - x1; 1 - x0.1; 2 - x0.01; 3 - x0.001; 4 - x0.0001; 5 - x10
1 1
C5 0
Main weight unit (Calibration and setting capacity weight units):
0 – kg,1 – lb
1 1
C6 6
Weight Units that can be chosen by UNIT key:
0-kg; 1-lb; 2-lb:oz; 3-kg/lb; 4-kg/lb:oz; 5-lb/lb:oz; 6-kg/lb/lb:oz;
(The limitation of units that can be used, refer to Table1 and Table2)
6 6
C7 9
Power-on zero-point range:
0 - calibration zero -point ±1%FS;
1 - calibration zero –point ±2%FS;
2 - calibration zero –point ±3%FS;
3 - calibration zero –point ±4%FS;
4 - calibration zero –point ±5%FS;
5 - calibration zero –point ±10%FS;
6 - calibration zero –point ±20%FS;
7 - calibration zero –point ±50%FS;
8 - calibration zero –point ±100%FS;
9 - no limitation;
9 9
PS102 www.measuretek.net .6