Datasheet
SPECIFICATION
MPT-45A MPT-45B MPT-45CMODEL
DC VOLTAGE
RATED CURRENT
CURRENT RANGE
RATED POWER
OUTPUT
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
VOLTAGE RANGE
FREQUENCY RANGE
EFFICIENCY(Typ.)
INPUT
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT
Note.7
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSIONOTHERS
NOTE
PACKING
OVERLOAD
OVER VOLTAGE
AC CURRENT (Typ.)
5V 12V
CH2
-5V
CH3
5V
CH1
12V
CH2
-12V
CH3
5V
CH1
15V
CH2
-15V
CH3
3A 2A 0.3A 3A 2A 0.3A 3A 1.6A 0.3A
0.4 ~ 5A 0.2 ~ 2.5A 0 ~ 0.5A 0.4 ~ 5A 0.2 ~ 2.5A 0 ~ 0.5A 0.4 ~ 5A 0.2 ~ 2.3A 0 ~ 0.5A
40.5W 42.6W 43.5W
52W with 18CFM min. Forced air convection
60mVp-p 120mVp-p 60mVp-p 60mVp-p 120mVp-p 100mVp-p 60mVp-p 120mVp-p 100mVp-p
±4.0% ±4.0% ±4.0%
CH1:4.5 ~ 5.5V
±7.0% ±5.0% ±5.0% ±5.0%±7.0% ±7.0%
±1.0% ±1.0% ±1.0%±2.0% ±1.0% ±1.0% ±1.0%
±1.0% ±1.0% ±1.0%
±2.0% ±2.0%
±3.0% ±3.0% ±3.0%±4.0% ±4.0% ±4.0%
90 ~ 264VAC 127 ~ 370VDC
47 ~ 440Hz
73% 75% 75%
1.2A/115VAC 0.7A/230VAC
COLD START 20A/115VAC 40A/230VAC
53 ~ 75W rated output power
5.75 ~ 6.75VDC on CH1
Protection type : Hiccup mode, recovers automatically after fault condition is removed
Protection type : Hiccup mode, recovers automatically after fault condition is removed
-20 ~ +85 , 10 ~ 95% RH℃
± ℃ ℃0.04%/ (0 ~ 50 ) on +5V output
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
271.5Khrs min. MIL-HDBK-217F (25 )℃
127*76*28mm (L*W*H)
0.21Kg; 72pcs/18Kg/1.35CUFT
File Name:MPT- 45 -SPEC 2014-01-29
800ms, 20ms/230VAC 800ms, 20ms/115VAC at full load
80ms/230VAC 12ms/115VAC at full load
-10 ~ +60 (Refer to "Derating Curve")℃
20 ~ 90% RH non-condensing
SAFETY STANDARDS
ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2, medical level, criteria A
VOLTAGE ADJ. RANGE
45W Triple Output Medical Type
MP T-45
s e ries
‧
Universal AC input / Full range
‧
Low leakage current <250 A
‧
Protections: Short circuit / Overload / Over voltage
‧
Cooling by free air convection
‧
Medical safety approved (2 x MOPP between primary to secondary)
‧
100% full load burn-in test
‧
Fixed switching frequency at 45KHz
‧
3 years warranty
■
Features :
RIPPLE & NOISE (max.)
Note.2
VOLTAGE TOLERANCE
Note.3
ENVIRONMENT
SAFETY &
EMC
(Note 4)
PROTECTION
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC 1min.
I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25 / 70% RH℃
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.℃
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. Mounting holes M1 and M2 should be grounded for EMI purposes.
6. Heat Sink HS1,HS2 can not be shorted.
7. .Touch current was measured from primary input to DC output
OUTPUT POWER (max.)
CH1
OUTPUT NUMBER
EMC IMMUNITY
EMC EMISSION
Earth leakage current < 250 A/264VAC , Touch current < 60 A/264VACμ
μ
μ