medicalManual.RevD.qxd 9/9/04 4:18 PM Page 2 DECLARATION OF CONFORMITY Manufacturer's Name: Maxtec®, Inc. Manufacturer's Address: Maxtec®, Inc. 6526 South Cottonwood Street Salt Lake City, Utah 84107 USA European Representative: Bio MS (Airox) L'Echangeur, Parc d'activite Pau-Pyrenees 64000 Pau, France Product: Oxygen Analyzer Model(s): MAXO2®+ A, MAXO2®+ AE Classification: IIa Classification criteria: Clause 3.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 4 Before use, all individuals who will be using the MAXO2® + must become thoroughly familiar with the information contained in this Operation Manual. Strict adherence to the operating instructions is necessary for safe, effective product performance. This product will perform only as designed if installed and operated in accordance with the manufacturer’s operating instructions. Use only genuine Maxtec® accessories and replacement parts.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 6 TABLE OF CONTENTS 1.0 SYSTEM OVERVIEW 1.0 1.1 Base Unit Description SYSTEM OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 1.1 1.2 1.3 1.4 2.0 Base Unit Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 Component Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 8 1.2 Component Identification 1.4 MAX-250+ Oxygen Sensor 1.3 Component Description The MAX-250+ oxygen sensor offers stability and extra life. The MAX-250+ is a galvanic, partial pressure sensor that is specific to oxygen. It consists of two electrodes (a cathode and an anode), a teflon membrane and an electrolyte. Oxygen diffuses through the teflon membrane and immediately reacts at a gold cathode.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 10 2.1.3 Automatic Calibration • If you are unsure about the displayed O2 percentage. (see Factors influencing accurate readings.) After the unit is turned on it will automatically calibrate to room air. The display should be stable and reading 20.9%. A simple calibration may be made with the sensor open to static Ambient air.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 12 4. Allow the oxygen to saturate the sensor. Although a stable value is usually observed within 30 seconds, allow at least two minutes to ensure that the sensor is completely saturated with the calibration gas. 5. If the MAXO2® + is not already turned on, do so now by pressing the analyzer “ON” button. 6. Press the Cal button on the MAXO2® + until you read the word CAL on the analyzer display. This can take approximately 3 seconds.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 14 CAL Err St: O2 Sensor reading not stable Wait for displayed oxygen reading to stabilize, when calibrating the device at 100% oxygen. Carefully, bring the two halves of the case together while positioning the wires so they are not pinched between the two case halves. The gasket separating the halves will be captured on the back case half.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 16 Reconnect the oxygen sensor to the connector on the printed circuit board. Carefully bring the two halves of the case together while positioning the wires to ensure they are not pinched between the two case halves. Make sure the sensor is fully inserted and in the proper orientation. Reinsert the three screws and tighten until the screws are snug. Verify the unit operates properly.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 18 BAT CAL Warm-up Time: Operating Temperature: Storage Temperature: Atmospheric Pressure Humidity: Power Requirements: Battery Life: Low Battery Indication: Sensor Type: Expected Sensor Life: A Model Dimensions: A Weight: AE Model Dimensions: AE Weight: 9.2 Standard Replacement Parts none required 15°C - 40°C (59°F - 104°F) -15°C - 50°C (5°F - 122°F) 800-1013 mBars 0-95% (non-condensing) 2, AA Alkaline batteries (2 x 1.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 20 10.0 WARRANTY The MAXO2® + Analyzer is designed for medical oxygen delivery equipment and systems. Under normal operating conditions, Maxtec® warrants the MAXO2® + Analyzer to be free from defects of workmanship or materials for a period of 2-years from the date of shipment from Maxtec®, provided that the unit is properly operated and maintained in accordance with Maxtec®’s operating instructions.