Datasheet

Table 18. Event Timing 2 Register
WRITE OPCODE
40h
READ OPCODE
C0h
FLASH STORED
Yes
FACTORY-STORED FLASH VALUE
0000h
Bit 15 14 13 12 11 10 9 8
Name TMM4 TMM3 TMM2 TMM1 TMM0 CAL_USE CAL_CFG2 CAL_CFG1
Bit 7 6 5 4 3 2 1 0
Name CAL_CFG0 TP1 TP0 PRECYC2 PRECYC1 PRECYC0 PORTCYC1 PORTCYC0
BIT NAME DESCRIPTION
15:11 TMM[4:0]
Temperature Measurements: These bits dene the number of temperature measurement cycles to
be executed when the EVTMG1 or EVTMG3 command is executed.
Cycles = 1+ TMM[4:0]
TMM[4:0] (decimal) CYCLES
0 1
1 2
…. ….
30 31
31 32
10 CAL_USE
Calibration Usage: This bit, when set, causes the MAX35101 to use the calibration data in the
CalibrationInt and CalibrationFrac registers during measurement, averaging and accumulation
of data while executing the EVTMG commands. All time measurements are scaled using the
calibration factors as described by the calibrate command.
9:7 CAL_CFG[2:0]
Calibration Conguration: These bits dene the point in the EVTMGx cycle/sequence where the
automatic calibration command is executed.
CAL_CFG[2:0]
DESCRIPTION
DURING EVTMGX SEQUENCES, AUTOMATIC EXECUTION OF
THE CALIBRATE COMMAND OCCURS AT:
000b through 011b Autocalibration disabled
100b
The beginning of each TOF_DIFF cycle
The beginning of each temperature cycle
101b
The beginning of each TOF_DIFF cycle
The beginning of each temperature sequence
110b
The beginning of each TOF_DIFF sequence
The beginning of each temperature cycle
111b
The beginning of each TOF_DIFF sequence
The beginning of each temperature sequence
MAX35101 Time-to-Digital Converter with Analog Front-End
www.maximintegrated.com
Maxim Integrated
47