Table 2. Configuration Register Bit Descriptions
Table 3. Resolution Configuration Table 5. Register Pointer Definition
Table 4. Fault Tolerance Configuration
F1 F0
CONSECUTIVE OUT-OF-LIMITS
CONVERSIONS TO TRIGGER OS
0 0 1
0 1 2
1 0 4
1 1 6
BIT NAME FUNCTION
0
Reserved
Power-up state = 0
The master can write to this bit, but it will always read out as a 0.
R1
Conversion Resolution Bit 1
Power-up state = 0
Sets conversion resolution (see Table 3)
R0
Conversion Resolution Bit 0
Power-up state = 0
Sets conversion resolution (see Table 3)
F1
Thermostat Fault Tolerance Bit 1
Power-up state = 0
Sets the thermostat fault tolerance (see Table 4).
F0
Thermostat Fault Tolerance Bit 0
Power-up state = 0
Sets the thermostat fault tolerance (see Table 4).
POL
Thermostat Output (OS) Polarity
Power-up state = 0
POL = 0 — OS is active low.
POL = 1 — OS is active high.
TM
Thermostat Operating Mode
Power-up state = 0
TM = 0 — Comparator mode
TM = 1 — Interrupt mode
See the Thermostat section for a detailed description of these modes.
SD
Shutdown
Power-up state = 0
SD = 0 — Active conversion and thermostat operation.
SD = 1 — Shutdown mode.
See the Shutdown Mode section for a detailed description of this mode.
R1 R0
THERMOMETER
RESOLUTION
MAX CONVERSION
TIME (ms)
0 0 9-bit 150
0 1 10-bit 300
1 0 11-bit 600
1 1 12-bit 1200
REGISTER P1 P0
Temperature 0 0
Conguration 0 1
T
HYST
1 0
T
OS
1 1
DS75LV Digital Thermometer and Thermostat