Datasheet
DS1921H/Z
35 of 45
Read Memory with CRC (invalid address)
RST PD Select RMC TA <00 to EOP> CRC16\ <32 bytes> CRC16\
The "32 bytes" are all 00h.
Clear Memory
RST PD Select CM FF loop
To verify success, read the Status Register at address 0214h. If MEMCLR is 1, the command was
executed successfully.
Convert Temperature
RST PD Select CT FF loop
To read the result and to verify success, read the addresses 0211h (result) and the Device Samples Coun-
ter at address 021Dh to 021Fh. If the count has incremented, the command was executed successfully.
Loop