Manual
71M6541 Demo Board REV 3.0 User’s Manual 
52    Rev 4.0 
Figure 2-12: Meter with Calibration System 
Figure 2-13 shows the screen on the controlling PC for a typical Demo Board. The error numbers are given in 
percent. This means that for the measured Demo Board, the sum of all errors resulting from tolerances of PCB 
components, current sensors, and 71M6541F tolerances was –3.41%, a range that can easily be compensated 
by calibration. 
Figure 2-14 shows a load-line obtained with a 71M6541 in differential mode. As can be seen, dynamic ranges of 
2,000:1 for current can be achieved with good circuit design, layout, cabling, and, of course, good current sen-
sors. 
Figure 2-13: Calibration System Screen 
Calibrator
AC Voltage
Current CT
Meter 
under 
Test
Optical Pickup 
for Pulses
Calibrated 
Outputs
Pulse 
Counter
PC